Open topic with navigation
Grid Interval Settings
The dialog is a part of the Grid Scan and Feature Scan wizards.
To set up parameters for grid scanning:
- In the Turn Mode you can control how the instrument turns to the grid points:
- Normal mode requires the instrument to turn to each point and measure it.
- Non-Stop mode allows the instrument to continuously turn and measure at the same time.
For IS / QS / GPT-9 Series, you can select either Normal or Non-Stop mode.
For DS-i, the Normal mode is always selected.
- In the Detail Mode select Coarse or Fine mode for the distance measurement during scanning procedure.
- For scanning procedure the "NP" is always selected in the NP Mode.
- Select the type of Interval for scanning. You can specify scanning interval using:
- Points - enter the number of scanning points on the horizontal axe (field H) and vertical axe (field V). After clicking the Calc button, the HA and VA fields display the corresponding horizontal and vertical angles interval for the given area.
- Angle - enter angles interval for in the HA and VA fields.
- Distance - enter the distance between neighboring points on the horizontal (HD Div) and vertical (VD Div) axes. Enter the distance from the robot to center of the selected area into the At Distance field, or click the Meas button to measure the distance value (for this measuring you can work without prism (NP button) / with prism (P button) / using Long Non-Prism mode (LNP button). After clicking the Calc button, the HA and VA fields display the corresponding angles interval for the given area.
- Click Next to start the scanning.
Open topic with navigation